Pullout strengths of orthodontic palatal mini-implants tested in vitro
-
The thickness of anterior palatal mucosa was designed as 3
mm and an Absoanchor (2.0
×
8
mm) was self-drill inserted into the Sawbones (cortical bone: 2
mm thickness; cancellous bone: 20
mm thickness) with The thickness of anterior palatal mucosa was designed as 3
mm and an Absoanchor (2.0
×
8
mm) was self-drill inserted into the Sawbones (cortical bone: 2
mm thickness; cancellous bone: 20
mm thickness) with 5
mm depth. -
An orthodontic wire (0.018 inch) was passed through the hole of the mini-implant and the n tied into the material testing machine (Lloyd). (A) Vertical pullout test; (B) horizontal pullout test.An orthodontic wire (0.018 inch) was passed through the hole of the mini-implant and the n tied into the material testing machine (Lloyd). (A) Vertical pullout test; (B) horizontal pullout test.