Journal of Dental Sciences
Volume 4, Issue 2 , Pages 75-80, June 2009

In vitro evaluation of the accuracy of Root ZX series electronic apex locators

  • Bor-Ren Duh

      Affiliations

    • Corresponding Author InformationCorresponding author. Department of Dentistry, Division of Endodontics, Mackay Memorial Hospital, 92, Chung-Shan North Road, Section 2, Taipei 10449, Taiwan

Department of Dentistry, Division of Endodontics, Mackay Memorial Hospital, Taipei, Taiwan

Received 7 February 2009; accepted 26 April 2009.

Article Outline

Background/Purpose

To evaluate the accuracy of Root ZX series electronic apex locators (EALs) for locating the apical constriction.

Materials and methods

Forty-two extracted human teeth were mounted in the same experimental apparatus used for research on Root ZX series EALs. Each root canal was measured with all Root ZX series EALs (5 groups). In Groups 1 (Root ZX) and 2 (DentaPort ZX), the apex was located with an EAL only; in Groups 3 (Solfy ZX), 4 (TriAuto ZX) and 5 (DentaPort ZX), the apex was located with an EAL and handpiece in the passive mode. The actual canal length was visually measured by inserting a size 15 K-file until its tip could be observed at the major apical foramen under 16 times magnification, and the working length was determined by subtracting 0.5 mm from this length. The experiment was set up, and the devices were used to detect the apical constriction when the meter value reached the “0.5” mark on each EAL.

Results

The results demonstrated that the mean distance between the file and apical constriction was 0.10–0.19 mm more apically located than those indicated with the Root ZX series EALs. Furthermore, the accuracy of the Root ZX series EALs in determining the working length within ± 0.5 mm from the apical constriction varied from 90.48% to 97.62%. No significant difference was found among the experimental groups (P > 0.05).

Conclusion

The results of the present study confirm that Root ZX series EALs can accurately determine the apical constriction.

KEY WORDS:  apex locator , DentaPort ZX , Root ZX , Solfy ZX , TriAuto ZX

No full text is available. To read the body of this article, please view the PDF online.

 

Back to Article Outline

References 

  1. Duh BR . Electronic root canal length measurement–a literature review . Chin Dent J . 2005;1:67–80
  2. Sunada I . New method for measuring the length of the root canal . J Dent Res . 1962;41:375–387
  3. Kuttler Y . Microscopic investigation of root apexes . J Am Dent Assoc . 1955;50:544–552
  4. Ricucci D , Langeland K . Apical limit of root canal instrument and obturation, part 2: a histological study . Int Endod J . 1998;31:394–409
  5. Sunada I . New method for measuring the length of the root canal . Jpn J Stomatol . 1958;25:161–171
  6. Kobayashi C , Okiji T , Kawashima N , Suda H , Sunada I . A basic study on the study on the electronic root canal length measurement, part 3: newly designed electronic root canal length measuring device using division method . Jpn J Conserv Dent . 1991;34:1442–1448
  7. Kobayashi C , Suda H . New electronic canal measuring device based on the ratio method . J Endod . 1994;20:111–114
  8. Kobayashi C , Matoba K , Suda H , Sunada I . New practical model of the division method electronic root canal length measuring device . J Jpn Endod Assoc . 1991;12:143–148
  9. Kobayashi C , Suda H . A basic study on the electronic root canal length measurement, part 4: a comparison of six apex locators . Jpn J Conserv Dent . 1993;36:185–192
  10. Kobayashi C . Electronic canal length measurement . Oral Surg Oral Med Oral Pathol . 1995;79:226–231
  11. Kobayashi C , Yoshioka T , Suda H . A new ultrasonic canal preparation system with electronic monitoring of file tip position . J Endod . 1996;22:489–492
  12. Kobayashi C , Matsumoto A , Nakajima M , Okiji T , Sunada I . Ultrasonic root canal preparation in the simultaneous use of the electronic canal length measuring device . Jpn J Conserv Dent . 1989;32:1198–1205
  13. Campbell D , Friedman S , Nguyen HQ , Kaufman A , Keila S . Apical extent of rotary canal instrumentation with an apexlocating handpiece in vitro . Oral Surg Oral Med Oral Pathol . 1998;85:319–324
  14. Jakobson SJM , Westphalen VPD , da Silva Neto UX , Fariniuk LF , Picoli F , Carneiro E . The accuracy in the control of the apical extent of rotary canal instrumentation using Root ZX II and ProTaper instruments: an in vivo study . J Endod . 2008;34:1342–1345
  15. Felippe WT , Felippe MCS , Carmona JR , Crozoé FCI , Alvisi BB . Ex vivo evaluation of the ability of the ROOT ZX II to locate the apical foramen and to control the apical extent of rotary canal instrumentation . Int Endod J . 2008;41:502–507
  16. Uzun O , Topuz O , Tinaz C , Nekoofar MH , Dummer PMH . Accuracy of two root canal length measurement devices integrated into rotary endodontic motors when removing gutta-percha from root-filled teeth . Int Endod J . 2008;41:725–732
  17. Kobayashi C , Suda H , Sunada I . A basic study on the electronic root canal length measurement, part 1: measurement on the extracted human tooth model . Jpn J Conserv Dent . 1991;34:1051–1055
  18. Kobayashi C , Yoshioka T , Suda H . A new engine-driven canal preparation system with electronic canal measuring capability . J Endod . 1997;23:751–754
  19. Kobayashi C , Suda H , Sunada I . A basic study on the electronic root canal length measurement, part 1: measurement using impedance analyzer . Jpn J Conserv Dent . 1991;34:1208–1221
  20. Oishi A , Yoshioka T , Kobayashi C , Suda H . Location of the file tip and the meter value of apex locators . Jpn J Conserv Dent . 2000;43:1145–1153
  21. Ushiyama J . New principle and method for measuring the root canal length . J Endod . 1983;9:97–104
  22. Venturi M , Breschi L . A comparison between two electronic apex locators: an ex vivo investigation . Int Endod J . 2007;40:362–373
  23. Meares WA , Steiman HR . The influence of sodium hypochlorite irrigation on the accuracy of the Root ZX electronic apex locator . J Endod . 2002;28:595–598
  24. Oishi A , Yoshioka T , Kobayashi C , Suda H . Electronic detection of root canal constrictions . J Endod . 2002;28:361–364
  25. Krajczár K , Marada G , Gyulai G , Tóth V . Comparison of radiographic and electronical working length determination on palatal and mesio-buccal root canals of extracted upper molars . Oral Surg Oral Med Oral Pathol Oral Radiol Endod . 2008;106:e90–e93
  26. Welk AR , Baumgartner JC , Marshall JG . An in vivo comparison of two frequency-based EALs . J Endod . 2003;29:497–500
  27. Vajrabhaya L , Tepmongkol P . Accuracy of apex locator . Endod Dent Traumatol . 1997;13:180–182
  28. Plotino G , Grande NM , Brigante L , Lesti B , Somma F . Ex vivo accuracy of three electronic apex locators: Root ZX, Elements Diagnostic Unit and Apex Locator and ProPex . Int Endod J . 2006;39:408–414
  29. Shabahang S , Goon WW , Gluskin AH . An in vivo evaluation of Root ZX electronic apex locator . J Endod . 1996;22:616–618
  30. Grimberg F , Banegas G , Chiacchio L , Zmener O . In vivo determination of root canal length: a preliminary report using the Tri Auto ZX apex-locating handpiece . Int Endod J . 2002;35:590–593
  31. D'Assuncao FLC , Albuquerque DS , Salazar-Silva JR , Queiroz Ferreira LC , Bezerra PM . The accuracy of root canal measurements using the Mini Apex Locator and Root ZX-II: an evaluation in vitro . Oral Surg Oral Med Oral Pathol Oral Radiol Endod . 2007;104:e50–e53
  32. Ebrahim AK , Wadachi R , Suda H . In vitro evaluation of the accuracy of five different electronic apex locators for determining the working length of endodontically retreated teeth . Aust Endod J . 2007;33:7–12
  33. Thomas AS , Hartwell GR , Moon PC . The accuracy of the Root ZX electronic apex locator using stainless-steel and nickel-titanium files . J Endod . 2003;29:662–663

PII: S1991-7902(09)60011-8

doi:10.1016/S1991-7902(09)60011-8

Journal of Dental Sciences
Volume 4, Issue 2 , Pages 75-80, June 2009