In vitro evaluation of the accuracy of Root ZX series electronic apex locators
Article Outline
Background/Purpose
To evaluate the accuracy of Root ZX series electronic apex locators (EALs) for locating the apical constriction.
Materials and methods
Forty-two extracted human teeth were mounted in the same experimental apparatus used for research on Root ZX series EALs. Each root canal was measured with all Root ZX series EALs (5 groups). In Groups 1 (Root ZX) and 2 (DentaPort ZX), the apex was located with an EAL only; in Groups 3 (Solfy ZX), 4 (TriAuto ZX) and 5 (DentaPort ZX), the apex was located with an EAL and handpiece in the passive mode. The actual canal length was visually measured by inserting a size 15 K-file until its tip could be observed at the major apical foramen under 16 times magnification, and the working length was determined by subtracting 0.5 mm from this length. The experiment was set up, and the devices were used to detect the apical constriction when the meter value reached the “0.5” mark on each EAL.
Results
The results demonstrated that the mean distance between the file and apical constriction was 0.10–0.19 mm more apically located than those indicated with the Root ZX series EALs. Furthermore, the accuracy of the Root ZX series EALs in determining the working length within ± 0.5 mm from the apical constriction varied from 90.48% to 97.62%. No significant difference was found among the experimental groups (P > 0.05).
Conclusion
The results of the present study confirm that Root ZX series EALs can accurately determine the apical constriction.
KEY WORDS: apex locator , DentaPort ZX , Root ZX , Solfy ZX , TriAuto ZX
No full text is available. To read the body of this article, please view the PDF online.
References
- . Electronic root canal length measurement–a literature review . Chin Dent J . 2005;1:67–80
- . New method for measuring the length of the root canal . J Dent Res . 1962;41:375–387
- . Microscopic investigation of root apexes . J Am Dent Assoc . 1955;50:544–552
- . Apical limit of root canal instrument and obturation, part 2: a histological study . Int Endod J . 1998;31:394–409
- . New method for measuring the length of the root canal . Jpn J Stomatol . 1958;25:161–171
- . A basic study on the study on the electronic root canal length measurement, part 3: newly designed electronic root canal length measuring device using division method . Jpn J Conserv Dent . 1991;34:1442–1448
- . New electronic canal measuring device based on the ratio method . J Endod . 1994;20:111–114
- . New practical model of the division method electronic root canal length measuring device . J Jpn Endod Assoc . 1991;12:143–148
- . A basic study on the electronic root canal length measurement, part 4: a comparison of six apex locators . Jpn J Conserv Dent . 1993;36:185–192
- . Electronic canal length measurement . Oral Surg Oral Med Oral Pathol . 1995;79:226–231
- . A new ultrasonic canal preparation system with electronic monitoring of file tip position . J Endod . 1996;22:489–492
- . Ultrasonic root canal preparation in the simultaneous use of the electronic canal length measuring device . Jpn J Conserv Dent . 1989;32:1198–1205
- . Apical extent of rotary canal instrumentation with an apexlocating handpiece in vitro . Oral Surg Oral Med Oral Pathol . 1998;85:319–324
- . The accuracy in the control of the apical extent of rotary canal instrumentation using Root ZX II and ProTaper instruments: an in vivo study . J Endod . 2008;34:1342–1345
- . Ex vivo evaluation of the ability of the ROOT ZX II to locate the apical foramen and to control the apical extent of rotary canal instrumentation . Int Endod J . 2008;41:502–507
- . Accuracy of two root canal length measurement devices integrated into rotary endodontic motors when removing gutta-percha from root-filled teeth . Int Endod J . 2008;41:725–732
- . A basic study on the electronic root canal length measurement, part 1: measurement on the extracted human tooth model . Jpn J Conserv Dent . 1991;34:1051–1055
- . A new engine-driven canal preparation system with electronic canal measuring capability . J Endod . 1997;23:751–754
- . A basic study on the electronic root canal length measurement, part 1: measurement using impedance analyzer . Jpn J Conserv Dent . 1991;34:1208–1221
- . Location of the file tip and the meter value of apex locators . Jpn J Conserv Dent . 2000;43:1145–1153
- . New principle and method for measuring the root canal length . J Endod . 1983;9:97–104
- . A comparison between two electronic apex locators: an ex vivo investigation . Int Endod J . 2007;40:362–373
- . The influence of sodium hypochlorite irrigation on the accuracy of the Root ZX electronic apex locator . J Endod . 2002;28:595–598
- . Electronic detection of root canal constrictions . J Endod . 2002;28:361–364
- . Comparison of radiographic and electronical working length determination on palatal and mesio-buccal root canals of extracted upper molars . Oral Surg Oral Med Oral Pathol Oral Radiol Endod . 2008;106:e90–e93
- . An in vivo comparison of two frequency-based EALs . J Endod . 2003;29:497–500
- . Accuracy of apex locator . Endod Dent Traumatol . 1997;13:180–182
- . Ex vivo accuracy of three electronic apex locators: Root ZX, Elements Diagnostic Unit and Apex Locator and ProPex . Int Endod J . 2006;39:408–414
- . An in vivo evaluation of Root ZX electronic apex locator . J Endod . 1996;22:616–618
- . In vivo determination of root canal length: a preliminary report using the Tri Auto ZX apex-locating handpiece . Int Endod J . 2002;35:590–593
- . The accuracy of root canal measurements using the Mini Apex Locator and Root ZX-II: an evaluation in vitro . Oral Surg Oral Med Oral Pathol Oral Radiol Endod . 2007;104:e50–e53
- . In vitro evaluation of the accuracy of five different electronic apex locators for determining the working length of endodontically retreated teeth . Aust Endod J . 2007;33:7–12
- . The accuracy of the Root ZX electronic apex locator using stainless-steel and nickel-titanium files . J Endod . 2003;29:662–663
PII: S1991-7902(09)60011-8
doi:10.1016/S1991-7902(09)60011-8
© 2009 Association for Dental Sciences of The Republic of China. Published by Elsevier Inc. All rights reserved.